Also called: Method verification, Analyser validation, MS ISO 15189
Verifying a New Haematology Analyser Before Patient Testing
The maker has proved the analyser can perform. Verification proves it performs in your lab, with your staff and your patients. A practical plan for clinics and accredited labs.
Last updated:
Validation or verification?
Validation — the manufacturer
The maker establishes the analyser’s performance claims — precision, linearity, carryover, flagging — through studies used for regulatory approval.
Verification — your lab
Before reporting patient results, the lab confirms it can achieve those claims in its own setting. MS ISO 15189 requires this for accredited labs, and it is good practice for any clinic.
Plan before you start
- Write a verification plan listing each study, the number of samples and the acceptance criteria.
- Take acceptance criteria from the maker’s claims or from recognised performance specifications, and fix them before you see the data.
- Use the reagent, control and calibrator lots you will use in routine work.
- Recognised protocols help: CLSI EP15 is commonly used for precision and trueness, and ICSH publishes guidance on evaluating blood cell analysers.
The core studies
| Study | What you do | What it shows |
|---|---|---|
| Repeatability | Run the same samples (low, normal and high) many times in one run | Short-term precision for each parameter |
| Within-lab precision | Run controls or samples over several days — EP15 uses 5 replicates on 5 days | Day-to-day variation under routine conditions |
| Method comparison | Run patient samples across the reportable range on the new analyser and the current method or reference lab | Bias against the results clinicians are used to |
| Carryover | Run a high sample three times, then a low sample three times | Whether a high result contaminates the next sample |
| Linearity and reportable range | Run a dilution series of a high sample, including low platelet and white-cell levels | Over what range results can be reported |
| Flagging | Compare analyser flags with blood films on abnormal samples | How often flags miss or over-call abnormal cells |
Carryover, step by step
Following the ICSH approach, run a sample with high counts three times (H1, H2, H3) and then a sample with low counts three times (L1, L2, L3). Carryover (%) = (L1 − L3) ÷ (H3 − L3) × 100. Compare the figure for WBC, RBC, HGB and PLT with the maker’s claim. Low carryover matters most where a very high count can be followed by a very low one, such as a leukaemia sample followed by a neutropenic patient.
Reference intervals and sample modes
- Verify any reference intervals you adopt from the maker or another lab with a small group of healthy local individuals before using them; establish your own if they do not fit.
- Verify each sample mode you will use — venous whole blood, pre-diluted capillary, body fluid — separately.
- Pay attention to low platelet counts, which drive dengue management.
- Check that results, units and flags arrive correctly in the LIS or clinic system.
Read next Validating the LIS link
Records and after go-live
- Keep the plan, raw data, statistics, conclusion and the sign-off by the person responsible for the lab.
- Repeat a reduced verification after relocation, a major repair or a software change, as your quality procedures specify.
- From day one, run internal QC and enrol in EQA so performance is monitored continuously.
- The supplier’s application specialist can help run the studies, but the lab owns the acceptance decision.
Read next Daily QC after verification
Related guides
Hematology Analyser supply near you
Delivery, installation and training across Malaysia from Batu Caves, Kota Kinabalu and Kuching.
- Hematology Analyser Kuala LumpurKuala Lumpur
- Hematology Analyser SelangorSelangor
- Hematology Analyser Petaling JayaSelangor
- Hematology Analyser Shah AlamSelangor
- Hematology Analyser KlangSelangor
- Hematology Analyser SerembanNegeri Sembilan
- Hematology Analyser MelakaMelaka
- Hematology Analyser Johor BahruJohor
Frequently asked questions
Does a GP clinic need to verify its haematology analyser?
How many patient samples are needed for a method comparison?
How is carryover calculated?
Does a refurbished analyser need verification?
Get a written quotation
Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation with delivery, installation and training spelled out.