Haematology Analyser Calibration: When, How and Who Does It
Controls tell you whether today’s results can be trusted; calibration is what makes them right in the first place. How the two differ, what a calibration involves, and when an analyser really needs one.
Calibrators, controls and verification do different jobs
All three use material with known values, which is why they are easily confused. The difference is what each is allowed to change. A calibrator adjusts the analyser; a control only checks it; verification proves the whole method works in your hands before patients are tested.
| Calibrator | Control | Method verification | |
|---|---|---|---|
| Purpose | Sets the analyser’s calibration factors so results agree with reference values | Checks that results stay within limits, run after run | Proves the analyser meets its claims in your lab before go-live |
| Error it deals with | Systematic error — a constant bias in one direction | Detects random and systematic error but corrects neither | Measures precision, bias, carryover and range at one point in time |
| Values | Lot-specific assigned values, traceable to reference methods | Lot-specific targets and acceptable ranges for your analyser | Acceptance criteria written into your plan |
| How often | At installation and when a trigger below applies | Every day before patients, and after maintenance or lot changes | Once at installation, then after major changes |
| Changes the analyser? | Yes | Never | No |
Read next Daily QC, Levey-Jennings and EQAVerifying a new analyser
What a calibration involves
Which parameters
Usually only the directly measured ones — typically WBC, RBC, HGB, PLT and MCV or HCT, depending on how the analyser measures red cells. MCH and MCHC are calculated from them and follow automatically. The white-cell differential is normally not calibrated by the user; the instructions for use (IFU) list what can be adjusted on your model.
The calibrator
Most clinics and laboratories use the maker’s calibrator, whose assigned values are set for that analyser and that lot. Some reference laboratories calibrate with fresh whole blood whose values are assigned by reference methods, such as the ICSH methods for haemoglobin and packed cell volume — demanding work that is rarely practical in a clinic.
The run
The analyser measures the calibrator several times, compares the mean with the assigned values and works out new calibration factors. Most software accepts them only if the replicates agree closely and the change stays within set limits.
Afterwards
Run every level of control before any patient sample. Record the old and new factors, the calibrator lot and who did the work, and sign off the result.
Check these before you calibrate
Calibrating an analyser that is dirty, contaminated or imprecise bakes the fault into every result that follows. Rule out the simple causes first.
- Daily and periodic maintenance is up to date, and the background (blank) count passes.
- Reagents are within their printed and opened expiry, and the lines are primed without bubbles.
- Precision is acceptable — repeated runs of one sample agree within the maker’s limits.
- Carryover is within the maker’s claim, so a high sample does not contaminate the next.
- The calibrator is in date, stored as labelled (usually 2–8 °C), brought to room temperature and mixed gently but thoroughly — never shaken.
- The assigned values entered or scanned into the analyser match the calibrator lot in your hand.
Read next Background counts and maintenanceMeasuring carryover
When to recalibrate — and when not to
| Situation | Recalibrate? | Notes |
|---|---|---|
| New analyser at installation | Yes, or confirm the factory calibration | Usually done by the supplier’s engineer or application specialist |
| The maker’s routine interval | Yes | Commonly every six months; follow the IFU for your model |
| After a major repair or part replacement | Usually | For example the haemoglobin measuring cell, the counting aperture, a pump or the main board |
| After the analyser is moved | Check, and recalibrate if the maker requires it | Run controls before patients either way |
| Persistent shift across several control levels | Only after troubleshooting | Try fresh reagent, fresh controls and cleaning first; recalibrate if the shift remains |
| One control out on one day | No | Rerun and investigate — recalibrating on a single failure adds error rather than removing it |
| A new control lot reads differently | No | Set targets for the new lot instead; the analyser has not changed |
Read next Westgard rules and QC shifts
Who calibrates, and what to ask the supplier
On many compact clinic analysers, calibration is done by the supplier’s engineer at installation and at preventive-maintenance visits, while clinic staff run the daily controls. Larger laboratories usually calibrate in-house under a written procedure. Either way, settle these points in writing before you buy.
- Who calibrates, how often, and whether calibration is included in the service contract.
- Whether calibrator material is supplied for your own use, with its shelf life and opened stability.
- Which parameters the user can adjust, and which only the engineer can.
- A calibration report after every visit, showing the factors before and after.
- How calibration is verified between calibrations — for example by running material of known value without adjusting anything.
- MDA registration for the calibrator and controls, as well as for the analyser and reagents.
Read next What a quotation should includeReagent storage and stability
Related guides
Hematology Analyser supply near you
Delivery, installation and training across Malaysia from Batu Caves, Kota Kinabalu and Kuching.
- Hematology Analyser Kuala LumpurKuala Lumpur
- Hematology Analyser SelangorSelangor
- Hematology Analyser Petaling JayaSelangor
- Hematology Analyser Shah AlamSelangor
- Hematology Analyser KlangSelangor
- Hematology Analyser SerembanNegeri Sembilan
- Hematology Analyser MelakaMelaka
- Hematology Analyser Johor BahruJohor
Frequently asked questions
What is the difference between a calibrator and a control?
How often should a haematology analyser be calibrated?
Should I recalibrate when a control fails?
Can a clinic calibrate its own FBC machine?