Haematology Controls and Calibrators: What to Order and How Much

Control blood and calibrators decide whether an FBC analyser can be trusted, and they have the shortest dating of anything on the bench. What to order for your analyser, how vial stability sets the size of each order, how to keep the cold chain to East Malaysia, and what to ask the supplier.

Also called Hematology controls, Hematology calibrator, CBC control blood, Hematology calibration and controls

Published:

Three different purchases

Controls, calibrators and EQA samples are all blood-like materials with known values, but you buy them for different jobs and on different schedules. How to run controls and when to calibrate are covered in our quality-control and calibration guides; this page is about what to order.

ItemWhat it is forHow often you buy it
Control bloodChecks the analyser every day before patients; never used to adjust itEvery few weeks, as each lot’s stability runs out
CalibratorAdjusts the analyser’s calibration factors at installation, at the maker’s interval and after major repairsOnly when a calibration is due — it has the shortest dating of all
EQA samplesUnknown samples from an external scheme, compared with other laboratoriesAn annual subscription, with samples sent in rounds through the year

Match the control to your analyser

Haematology controls are made from processed blood cells that behave like patient samples on the analyser. Their target values are assigned for each analyser model and each lot, and listed on the lot’s assay sheet. A control without values for your exact analyser cannot tell you whether it is working.

Levels

Most haematology controls come in three levels — low, normal and high — so you check the analyser near the decision points as well as in the middle. Order the levels your analyser maker recommends.

3-part or 5-part

A control assayed for a 3-part analyser gives targets for the three white-cell groups only. A 5-part analyser needs a control with assigned values for all five cell types, and reticulocytes or nucleated red cells if you report them.

Model and software

Check that your analyser model appears on the assay sheet, and ask whether the values change after a software update or between sample modes, such as whole blood and pre-diluted.

Haemoglobin meters

Point-of-care haemoglobin meters use their own liquid controls, made for specific meters, not FBC control blood.

Choosing a haemoglobin meter

Stability decides how much to order

Haematology controls are made with real cells, so they expire quickly. Each one has two limits: closed-vial stability, from manufacture until you open it, and open-vial stability, from opening until you discard it. These are examples from one maker’s published range; check the insert for the product you are offered.

  • Closed-vial dating runs from manufacture, so a lot often reaches Malaysia with weeks already used. Agree a minimum remaining shelf life at delivery.
  • You cannot stock up: with a 75-day closed-vial life, more than about two months’ supply is likely to expire on the shelf.
  • With 7-day open-vial stability and daily QC, you open a new set roughly every week — four or five sets a month.
  • With 30-day open-vial stability, one set a month is usually enough, provided each vial holds enough for all your runs.
  • Check the volume each run uses against the vial fill volume, especially if you run controls more than once a day.
  • Order calibrator only when a calibration is booked: with five days open and 45 days closed, it cannot sit in the fridge waiting.
  • Overlap lots: run a new control lot alongside the old one for several days to set its targets before the old lot runs out.
Published stability, Streck haematology controls and calibrator (stored at 2–8 °C)
Product (use)Open vialClosed vial
Para 12 Extend (3-part analyser control)30 days190 days
Para 12 Plus (5-part control for a specific analyser)7 days75 days
Retic-Chex II (reticulocyte control)14 days75 days
HQ-Chex (haemoglobin and glucose meter control)30 days180 days
Cal-Chex (calibrator)5 days45 days

Keeping the cold chain — including to Sabah and Sarawak

Most haematology controls and calibrators are stored at 2–8 °C, and most labels say not to freeze them. Heat or freezing damages the cells, and the damage shows up as a QC failure that looks like an analyser fault.

  • Ask how the supplier ships: an insulated box with cold packs, and a temperature indicator or logger for longer journeys.
  • Air freight to East Malaysia adds transit time. Order earlier, and avoid deliveries that sit in a depot over a weekend or public holiday.
  • On arrival, check the temperature indicator, the lot, the expiry date and the appearance described in the insert. Record it, and refuse a delivery that failed the cold chain.
  • Store controls in a monitored fridge, never in the door, and not with food.
  • Bring vials to room temperature and mix them exactly as the insert says before use. Most say mix gently by inversion, not by shaking.
  • Write the opening date on every vial, and discard it at the end of its open-vial life even if some is left.

Manufacturer or third-party controls, and EQA

For external quality assessment, the Department of Standards Malaysia’s specific criteria for medical testing (SC 2) set an order of preference: a scheme accredited to ISO/IEC 17043 first, then a recognised scheme, then comparisons between accredited laboratories, and only then comparisons between non-accredited laboratories. Malaysian hospital laboratories report enrolling in RCPAQAP, the Royal College of Pathologists of Australasia programme, for full blood count and related haematology surveys.

Manufacturer controls

Made for the maker’s own analysers, and sometimes required by a reagent rental or warranty agreement. Read the contract before changing.

Reagent rental terms that tie you to one supplier

Third-party controls

Independent controls assayed for many analysers. MS ISO 15189 asks laboratories to consider independent control material, either instead of or in addition to the manufacturer’s, because it can expose a problem the matched reagent and control may share.

Peer comparison

Some control makers run peer-group programmes that compare your QC results with other laboratories using the same lot and analyser.

  • Ask each EQA scheme which parameters it covers — FBC, differential, reticulocytes — and how often samples are sent.
  • Check that samples can be shipped to your state, and how long they take.
  • Ask whether the scheme is accredited to ISO/IEC 17043.
  • Even a GP clinic can enrol. It is good practice for any site that reports results, and required for accreditation.

Questions to ask the supplier

  • Which control and calibrator products are assayed for our exact analyser model and software version?
  • What are the open- and closed-vial stabilities, and what minimum remaining shelf life will each delivery have?
  • Can you deliver on a standing monthly schedule, and how do you ship to our state?
  • What happens if a delivery arrives warm or late — replacement, and at whose cost?
  • Are controls and calibrators inside a reagent rental or service contract, or priced separately?
  • Who calibrates the analyser, and is the calibrator supplied for our own use?
  • Are the controls and calibrators registered with the MDA? Ask for the numbers and check them on the public register.

Frequently asked questions

How many levels of haematology control do I need?
Most analysers use three levels — low, normal and high. Order the levels your analyser maker recommends, with assigned values for your exact model.
How long does an opened haematology control last?
It depends on the product — from about a week to about a month. One maker lists 30 days for a 3-part control and 7 days for a 5-part control. Check the insert and date every vial when you open it.
Can haematology controls be frozen?
No. Most are stored at 2–8 °C and labelled do not freeze, because freezing damages the cells. Keep them in a monitored fridge, not in the door.
Can I use a control from another brand on my analyser?
Yes, if it is assayed for your analyser model, registered with the MDA and allowed by your contract. Independent controls are encouraged under MS ISO 15189, but some reagent rental agreements require the maker’s controls.
How often do I need to buy calibrator?
Only when a calibration is due — at installation, at the maker’s interval (commonly every six months) and after major repairs. Calibrator has very short stability, so order it for a booked date.
Does a GP clinic need EQA?
It is good practice for any site reporting FBC results, and essential for accreditation. Ask EQA schemes about FBC surveys, shipping to your state and ISO/IEC 17043 accreditation.

Get a written quotation

Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation that states who handles delivery, installation and training, and what each costs.