How to Read a Haematology Analyser Spec Sheet
Every brochure promises speed and precision. Line by line, what the numbers on a haematology analyser specification sheet mean, which ones matter for a clinic, and the questions that turn two brochures into a fair comparison.
Parameters and the differential
Reportable or research-only
Count the parameters you can put on a patient report. Some sheets include research-use-only (RUO) parameters and histograms in the headline number.
3-part or 5-part
A 3-part analyser sizes white cells by impedance into lymphocytes, mid-sized cells and granulocytes. A 5-part analyser separates neutrophils, lymphocytes, monocytes, eosinophils and basophils, usually by flow cytometry.
3-part vs 5-partExtra channels
Reticulocytes, nucleated red cells, immature granulocytes or CRP appear on some models. Check whether each needs its own reagent and whether it is reportable.
Reticulocyte countsHaemoglobin reagent
Check whether the haemoglobin method is cyanide-free. Cyanide-containing reagents make waste handling harder.
Throughput and sample modes
| Spec line | What it means | What to ask |
|---|---|---|
| Throughput (samples an hour) | The maximum rate, usually in the fastest mode | The rate for CBC with differential, in pre-diluted mode and with the autoloader, if fitted |
| Whole-blood sample volume | Blood aspirated per test — often about 10–20 µL on benchtop analysers | The minimum volume that must be in the tube for a reliable aspiration |
| Pre-diluted or capillary mode | A small exact volume diluted before the run | The volume needed, who does the dilution, and whether the mode is verified |
| Open tube, closed tube or autoloader | Manual aspiration from an open tube; piercing the cap; a rack sampler | Closed-tube sampling and an autoloader matter in busy laboratories |
| Start-up and shutdown | Time before the first result, and the cleaning cycle at the end of the day | Reagent used by each start-up and shutdown |
Read next Capillary and pre-diluted samples
Precision, linearity and carryover
These lines tell you how good the numbers are, and each means little without the conditions printed beside it.
| Spec | Typical wording | How to read it |
|---|---|---|
| Precision (repeatability) | WBC CV ≤2.5% (3.5–15.0 × 10⁹/L); PLT CV ≤4% (150–500 × 10⁹/L) | Lower is better. Check the level it applies to — precision at low platelet and white-cell counts matters most in dengue |
| Linearity | PLT 0–1,000 × 10⁹/L or higher | The range reported without dilution; above it, dilute and rerun |
| Carryover | WBC, RBC, HGB ≤0.5%; PLT ≤1% | How much of the previous sample contaminates the next — matters after a very high count |
| Comparison | Correlation coefficient against a reference analyser | Ask for the data and the comparator; a high correlation can still hide a bias |
Reagents and running cost
- Number of reagents: 3-part analysers often use a diluent, a lyse and a cleaner; 5-part analysers use more, sometimes including stains or dyes.
- Stability after opening: at low volume, a reagent can expire on board before it is used up.
- Reagent used per test and per start-up and shutdown — the fixed daily amount drives cost at low volume.
- Closed or open reagents: whether only the maker’s reagents may be used.
- Storage: controls and some reagents need a refrigerator; check the space and the temperature log.
- Liquid waste per test, and how it is collected and disposed of.
Data, connectivity and QC software
Storage
How many results the analyser stores with their histograms and scattergrams, and how they are backed up or exported.
LIS and barcode
Two-way connection to a laboratory or clinic system, usually by ASTM or HL7, and a barcode reader for sample identification.
LIS connectivityQC tools
Built-in Levey-Jennings charts, a moving-average check on patient results (X-B), user log-ins and an audit trail.
Daily QCSite needs and the regulatory line
- Footprint and weight, plus space beside it for reagent boxes and a waste container.
- Operating temperature and humidity range — usually an air-conditioned room.
- Power supply and consumption; a UPS stops a power cut from aborting a run mid-sample.
- MDA registration numbers for the analyser and its reagents; CE or FDA marks are foreign approvals and do not replace Malaysian registration.
- Who installs, trains and services it — the quotation should say so in writing.
Read next MDA registration explained
Related guides
Hematology Analyser supply near you
Written quotes for clinics across Malaysia, handled from Batu Caves, Kota Kinabalu and Kuching.
- Hematology Analyser Kuala LumpurKuala Lumpur
- Hematology Analyser SelangorSelangor
- Hematology Analyser Petaling JayaSelangor
- Hematology Analyser Shah AlamSelangor
- Hematology Analyser KlangSelangor
- Hematology Analyser SerembanNegeri Sembilan
- Hematology Analyser MelakaMelaka
- Hematology Analyser Johor BahruJohor
Frequently asked questions
What does CV mean on a haematology analyser spec sheet?
Is higher throughput always better?
What is carryover on an analyser spec sheet?
Which spec matters most for a GP clinic?
Get a written quotation