Skip to main content

AA Alive Sdn Bhd · SSM 1204108-DEst. 2016Superbrands Malaysia 2025Batu Caves · Kota Kinabalu · Kuching

WhatsApp 010-905 2891

How a Haematology Analyser Works: From Tube to FBC Report

What happens inside an FBC machine in the minute after you load a tube: how cells are counted and sized, how a 5-part analyser tells white cells apart, how haemoglobin is measured, and which numbers on the report are calculated rather than measured.

Also called How a hematology analyzer works, Hematology analyzer principle, FBC machine principle, Coulter principle

Published:

One sample, up to three measuring channels

The analyser aspirates a few microlitres of well-mixed EDTA blood and splits it. One portion is diluted heavily in an isotonic diluent so red cells and platelets can be counted. Another is mixed with a lyse reagent that destroys the red cells, so white cells can be counted and the freed haemoglobin measured. A 5-part analyser adds an optical channel in which a laser classifies each white cell. In one widely used design, blood for the red-cell and platelet count is diluted about 6,000-fold and blood for the white-cell count about 250-fold.

A typical layout. Channel names, dilutions and reagents differ by maker.
ChannelReagent addedWhat is measuredMethod
RBC / PLTDiluent onlyRed cells, platelets and their volumesImpedance (Coulter principle)
WBC / HGBDiluent and lyseWhite-cell count, 3-part sizing and haemoglobinImpedance for cells; photometry for haemoglobin
Optical (5-part analysers)Lyse plus a stain or fluorescent dyeFive white-cell types; on some models NRBC, reticulocytes and immature granulocytesLaser light scatter and fluorescence (flow cytometry)

The impedance (Coulter) principle

Wallace Coulter’s method, patented in 1953 and still the basis of most FBC machines, rests on one fact: blood cells are poor electrical conductors. Suspend them in a salt solution, draw it through a tiny aperture with an electrode on each side, and every cell that passes briefly raises the resistance of the aperture. Under a constant current that shows up as a voltage pulse. The number of pulses is the cell count; the height of each pulse is proportional to that cell’s volume.

Diagram of an impedance counting aperture: cells in single file pass through a small hole between two electrodes, each producing a voltage pulse; pulse heights are sorted into a size histogram with platelets below a discriminator and red cells above it.
Each cell that crosses the aperture produces one pulse. The analyser counts the pulses and sorts them by height into a size histogram, with a discriminator separating platelets from red cells.

Counting

Pulses are counted in a fixed volume of diluted sample and converted back to cells per litre of blood using the dilution factor.

Sizing

Pulse heights are sorted into a size histogram in femtolitres (fL). Red cells cluster around 80–96 fL in healthy adults and platelets far below; a discriminator between the two decides which pulses are counted as which.

Reading the histograms

Coincidence

Two cells crossing the aperture together make one large pulse. Heavy dilution and hydrodynamic focusing keep this rare, and the analyser corrects for what remains.

From Model A to the clinic analyser

Coulter’s Model A was the first commercial haematology analyser. By 1977 the Model S Plus added a three-part white-cell differential by size, the approach compact 3-part clinic analysers still use.

Hydrodynamic focusing: one cell at a time, through the centre

A cell that drifts through the edge of the aperture produces a distorted pulse and can be sized wrongly. Modern analysers inject the sample through a fine nozzle into a stream of sheath fluid, which narrows it into a single-file thread through the centre of the aperture. The FDA’s decision summary for Sysmex’s compact XW-100, for example, describes direct current with hydrodynamic focusing for every parameter except haemoglobin, which is measured photometrically.

  • Fewer coincidence events, so counts stay accurate at high cell numbers.
  • More consistent pulse heights, so a tighter, more reliable MCV and RDW.
  • Cells pass the sensing zone once, so they are not counted twice.

Optical flow cytometry: how a 5-part analyser sorts white cells

Size alone separates white cells into three groups. To report neutrophils, lymphocytes, monocytes, eosinophils and basophils separately, a 5-part analyser passes each white cell through a laser beam and measures the light it scatters. Forward scatter rises with cell size. Side scatter, measured at about 90°, rises with internal complexity such as granules and a lobed nucleus. Where a fluorescent dye is added, side fluorescence shows how much nucleic acid the cell holds. Sysmex’s XN-series, for example, reads forward scatter, side scatter and side fluorescence from a 633 nm laser.

Diagram of a laser crossing a flow cell with forward scatter, side scatter and fluorescence detectors, beside a schematic scattergram in which lymphocytes, monocytes, neutrophils and eosinophils form separate clusters.
Each white cell scatters laser light according to its size and granularity, and a fluorescent dye adds nucleic-acid content. Plotted together, the five cell types fall into separate clusters. Positions are schematic and differ by maker.

Other makers, other signals

Not every 5-part analyser uses fluorescence. Beckman Coulter’s DxH analysers, for instance, combine cell volume, conductivity (opacity) and light scatter for each cell.

When clusters run together

Abnormal cells such as blasts or immature granulocytes fall between or outside the normal clusters. The analyser cannot name them, so it flags the result for a blood film.

What the flags mean
CellWhere it fallsWhy
LymphocytesLow side scatterSmall cells with few granules
MonocytesModerate scatter, high fluorescenceLarger cells with a folded nucleus and more nucleic acid than neutrophils
NeutrophilsHigh side scatterGranular cytoplasm and a lobed nucleus
EosinophilsHighest side scatterLarge, dense granules
Basophils and immature cellsA further channel or region, depending on the makerThey overlap other cells in the main plot

Haemoglobin: photometry, now without cyanide

The lyse reagent breaks open the red cells, and the freed haemoglobin is converted to a single stable coloured compound whose absorbance a photometer reads. The reference method converts it to cyanmethaemoglobin, but cyanide waste is a hazard, so most analysers now use cyanide-free reagents. The best known, sodium lauryl sulfate (SLS) haemoglobin, was published by Oshiro and Takenaka in 1982 and is used by several analyser makers.

  • A blank reading of the reagents is taken in each cycle, so the analyser measures only what the blood adds.
  • Anything that makes the solution cloudy, such as lipaemia or a very high white-cell count, can raise the haemoglobin reading, and with it MCH and MCHC.
  • Because the lyse is shared, too little of it leaves red cells unlysed, which raises both the white-cell count and the haemoglobin.

Measured or calculated? Reading the report with that in mind

Only some lines on an FBC report are measured directly. The rest are calculated from them, so an error in one measured value spreads into every index built on it. That is why an impossible MCHC is such a useful warning sign.

Formulas as Boule gives them for its 3-part analyser. Some analysers derive HCT from the summed red-cell pulses instead.
ParameterHow it is obtainedFormula or source
WBC, RBC, PLTMeasured (counted)Impedance pulses, or optical events
HGBMeasuredPhotometry after lysis
MCVMeasured on impedance analysersMean volume of all counted red-cell pulses
HCTCalculatedRBC × MCV
MCHCalculatedHGB ÷ RBC
MCHCCalculatedHGB ÷ HCT
RDWFrom the RBC histogramSpread of red-cell volumes
MPV, PDWFrom the PLT histogramMean and spread of platelet volumes
PCT (plateletcrit)CalculatedPLT × MPV
Differential (% and #)Measured (classified)Size on a 3-part; scatter and fluorescence on a 5-part

What is optimal: sample age, mixing and QC timing

The analyser is only as good as the tube it is given and the controls run before it. These habits give the most reliable result.

StepOptimal practiceWhy
TubeK2 or K3 EDTA, filled to the lineAn under-filled tube leaves excess EDTA, which shrinks red cells and lowers HCT and MCV
Mixing at collectionInvert gently 8–10 times straight awayThe count BD gives for EDTA tubes; small clots ruin the platelet count
Mixing before analysisRe-mix by gentle inversion just before aspirationRed cells settle on standing; an unmixed sample reads falsely low or high
Sample ageRun within a few hours; for a differential, within 8 hours at room temperatureBoule states its M32 needs samples under 8 hours old for a correct white-cell differential
If there is a delayRefrigerate at 4 °C and run within 12 hoursIn a 48-hour study, differential counts changed significantly after 12 hours at 23 °C, while 4 °C kept the full count and differential stable for 12 hours
ControlsTwo levels before patient samples each testing day, and after every reagent lot change, maintenance or serviceCatches drift before it reaches a patient result; follow your maker’s interval where it is stricter
BackgroundA blank at start-up, within the maker’s limitA high background adds false counts, most visibly to platelets

Frequently asked questions

What principle does a haematology analyser use?
Most use the impedance (Coulter) principle to count and size red cells, platelets and, on 3-part models, white cells; photometry for haemoglobin; and, on 5-part models, laser light scatter and fluorescence to classify white cells.
Is the haematocrit measured or calculated?
On most impedance analysers it is calculated from the red-cell count and MCV (HCT = RBC × MCV), or summed from the red-cell pulses, rather than measured by spinning the blood as a manual microhaematocrit does.
Why does a 5-part analyser need a laser?
Because size alone cannot reliably separate eosinophils, basophils and monocytes. Light scatter adds information about granules and the nucleus, and fluorescent dyes add nucleic-acid content.
How long after collection can an FBC be run?
As soon as practical. RBC, HGB and platelet counts stay stable for longer, but the white-cell differential changes within hours at room temperature. If there is a delay, refrigerate the tube at 4 °C and follow your analyser maker’s stability limits.
Who invented the blood cell counter?
Wallace H. Coulter. His aperture-impedance method, the Coulter principle, was patented in 1953 and became the basis of the first commercial haematology analysers.

Get a written quotation

Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation that states who handles delivery, installation and training, and what each costs.

010-905 2891 · Mon–Sat, 9:00 am – 6:00 pm