Skip to main content

AA Alive Sdn Bhd · SSM 1204108-DEst. 2016Superbrands Malaysia 2025Batu Caves · Kota Kinabalu · Kuching

WhatsApp 010-905 2891

Haematology Analyser Installation: A Buyer’s Checklist from Site Preparation to Sign-Off

Most installation delays come from the room, not the analyser: no dedicated socket, no space behind it, no plan for the waste. What to prepare before delivery, what to check on the day, and the records to have in hand before the first patient result.

Also called Hematology analyzer installation, Site preparation, FBC machine installation requirements, Acceptance testing

Published:

Get the site requirements in writing before you order

Every analyser’s instructions for use set out its installation requirements, and the installer should send you a pre-installation checklist before delivery. The written quotation states who installs, calibrates and trains. As an example of what the instructions contain, Sysmex’s XN-L general information asks for an ambient temperature of 15–35 °C and 20–85% relative humidity, with air-conditioning if the room falls outside that; an earthed, dedicated power outlet within reach of its 2-metre cable; a bench or floor that can take the weight (about 35 kg for the XN-330); and at least 30 cm clear behind the analyser. A compact 3-part analyser is far smaller — Mindray describes the BC-30s footprint as similar to a 17-inch laptop — but the same questions apply.

Top-down plan of a laboratory bench: a haematology analyser against the wall with a hatched clear gap behind it, reagent containers on its left feeding it, a UPS wired to a dedicated earthed socket and a waste container on its right, a printer and a network line to a LAN point for the LIS, and a sample tray and sharps bin at the front. Notes below cover room climate, power and placement.
Plan the bench around the analyser’s service needs: clearance behind it, reagents and waste within tubing reach, a dedicated earthed socket through a UPS, and a network point for the LIS. Take the exact distances from your model’s instructions.

Site preparation checklist

ItemWhat to prepareWhy it matters
BenchStable, level, rated for the analyser’s weight; away from direct sun, heat sources and vibrationMovement and heat upset counting and the optics
ClearanceThe gap the instructions specify behind and beside the analyserVentilation, tubing and the engineer’s access for service
PowerAn earthed, dedicated socket within cable reach; no extension leads or multi-plugsShared or poorly earthed circuits cause errors and are a safety risk
UPSSized for the analyser’s rating (the XN-330 lists 235 VA) and long enough to finish a run and shut downA power cut mid-run can lose results and leave reagent in the lines
Room climateWithin the instructions’ range day and night, including weekends when air-conditioning may be offOut-of-range temperature affects both the analyser and its reagents
ReagentsShelf space beside or below the analyser within tubing reach, plus storage for stock at the labelled temperatureTubing length is fixed; stock must stay within its storage range
WasteA waste container or approved drain for the analyser’s liquid waste, handled under your clinical-waste arrangementsLiquid waste is potentially infectious and must not overflow onto the bench
NetworkA LAN point or agreed connection for the LIS, and space for a barcode reader and printerResults should reach the patient record without retyping

On installation day

  • Inspect the packaging on arrival and report any damage to the supplier before installation starts.
  • Check that the serial number and model match the quotation and the MDA registration.
  • The installer connects reagents and waste, runs the start-up self-checks and confirms the background count passes.
  • Calibration is performed, or the factory calibration confirmed, with the calibrator lot recorded.
  • Controls are run at every level and are within range before anything else.
  • If you connect to an LIS, send test results through and check patient ID, units, flags and reference ranges in the clinic system.
  • Each operator is trained on start-up, shutdown, daily and weekly maintenance, QC, error messages and sample handling — and signs a training record.

Acceptance and verification before the first patient

The WHO laboratory quality handbook says equipment should not be used until it is fully installed, its performance is verified and the staff testing on it are trained. The Department of Standards Malaysia’s criteria for accredited medical laboratories (SC 2) ask for commissioning records for a testing analyser before it is placed into use, and user acceptance testing for LIS links and software.

  • Precision: repeated runs of the same samples agree within the maker’s claims.
  • Comparison: your patient samples agree with the method or reference laboratory you used before.
  • Carryover: a high sample does not contaminate the next one.
  • Sample modes: each mode you will use — venous whole blood, pre-diluted capillary — is checked separately.
  • Sign-off: the person responsible for the laboratory accepts the analyser in writing before patient results are released.

Documents to receive and keep

  • The installation or commissioning report, signed and dated.
  • The calibration report and the first QC printouts.
  • Training records for every operator.
  • MDA registration numbers for the analyser and each reagent, control and calibrator, and a copy of the supplier’s establishment licence.
  • The warranty start date, the preventive-maintenance schedule and the service contact for your state.
  • The instructions for use, quick-reference guides and safety data sheets for the reagents.
  • Your verification plan, data and sign-off.

Moving or reinstalling later

Treat a move like a small installation. Ask the supplier whether the engineer must do it, check the new bench, power and climate against the same list, and run controls before patients. Recalibrate if the maker requires it after relocation, and repeat a reduced verification as your quality procedures specify.

Frequently asked questions

How much space does a haematology analyser need?
It depends on the model. A compact 3-part analyser takes about the bench space of a laptop; a small-lab 5-part analyser such as the XN-330 is about 45 cm wide and 46 cm deep, and its instructions ask for 30 cm clear behind it. Add room for reagents, waste and a printer.
Do I need a UPS for an FBC machine?
It is strongly advisable. Size it for the analyser’s power rating and long enough to finish a run and shut down safely, and ask the installer to confirm the model.
What room temperature does a haematology analyser need?
Follow the instructions for your model. As one example, Sysmex specifies 15–35 °C and 20–85% humidity for its XN-L analysers. Keep the room in range at night and at weekends too.
Who installs the analyser?
The written quotation states who installs, calibrates and trains, and what each costs. Usually it is the supplier’s engineer or an authorised service partner.
When can we start testing patients?
Only after installation, calibration, passing controls, operator training and your verification are complete and signed off.
Can we move the analyser ourselves?
Ask the supplier first; some warranties require the engineer to move it. Either way, run controls before patients, and recalibrate if the maker requires it.

Get a written quotation

Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation that states who handles delivery, installation and training, and what each costs.

010-905 2891 · Mon–Sat, 9:00 am – 6:00 pm